Keysight Technologies N9040B UXA Signal Analyzer Service Guide 183
Front End Control Troubleshooting
A15 Front End Control Assembly Troubleshooting
Verifying Option ALV
Equipment needed:
Volt meter
Adapters and cables to connect the volt meter to the Aux IF Out SMA
connector on the spectrum analyzer rear panel.
Signal source
Signal source settings:
Frequency: 1 GHz (CW)
Amplitude: 10 dBm
Press the following keys on the analyzer:
MODE/MEAS, Spectrum Analyzer, Mode Preset, OK, FREQ Channel, and set
the center frequency to 1 GHz. Press Span and toggle the switch to Zero Span.
Press Input/Output, Output, Aux IF Out and choose Fast Log Video from the
drop-down menu.
Connect the rear panel Aux IF Out of the signal analyzer to a voltmeter set to
measure DC volts.
Verify that signal level appears on the analyzer screen positioned 1 division
below top graticule line.
Voltmeter should measure approximately 1.5 Vdc.
Reduce source output power to 20 dBm. Signal should now appear one
division lower, and voltmeter should measure approximately 1.3 VDC.
Reduce source output power to 30 dBm. Signal should now appear one
division lower, and voltmeter should measure approximately 1.0 VDC.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9040B Option 508 and is the answer not in the manual?

Keysight N9040B Option 508 Specifications

General IconGeneral
BrandKeysight
ModelN9040B Option 508
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals