Keysight Technologies N9040B UXA Signal Analyzer Service Guide 197
Analog/Digital IF Troubleshooting
A3 Digital IF Assembly Description
A3 Digital IF Assembly Description
The A3 Digital I.F. has circuitry that is needed to analyze complex
communication signals that can occupy up to 40 MHz of information
bandwidth. This assembly has two inputs:
1. The final 22.5 MHz I.F. from the A2 Analog I.F. for IF bandwidth paths up
to 25 MHz wide. This input is used for swept and FFT spectrum analysis.
2. The 250 MHz IF from the A15 Front End Control Assembly for IF
bandwidth of 40 MHz.
Each input has a separate ADC that processes the time domain continuous
data into I/Q (in-phase and quadrature) signals before sending the data to the
A4 CPU assembly for further processing and front panel display.
The Digital I.F. has the following main functions:
Digitizes the 22.5 MHz and the 250 MHz IF signals
16 bit ADC with a 100 MHz sample rate to digitize the 22.5 MHz IF
12 bit ADC with a 200 MHz sample rate to digitize the 250 MHz IF
DSP for standard spectrum analysis
Capture memory for complex signals
—Noise Source Control
Dither for final I.F
Provides the alignment sequence generator for wide band alignments
Provides wide band Comb Calibration Signal
Trigger interpolation and associated alignment

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9040B Option 508 and is the answer not in the manual?

Keysight N9040B Option 508 Specifications

General IconGeneral
BrandKeysight
ModelN9040B Option 508
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals