454 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Assembly Replacement Procedures
Fan Assembly
Replacement
1. Refer to Figure 15-100. Place the fan bracket next to the chassis and plug
in the fan wires into the rear mother board.
2. For Serial Number Prefix < 5605 Options 508, 513, 526:
Refer to Figure 15-98. Place the fan bracket into position taking care
to avoid pinching the fan wires between the fan bracket and the
chassis. Replace the nine screws (1) (0515-0372) and the two
screws (2) (0515-1946). Torque all screws to 9 inch-pounds.
3. For Serial Number Prefix 5605 and All Options 544 and 550:
Refer to Figure 15-99. Place the fan bracket into position taking care
to avoid pinching the fan wires between the fan bracket and the
chassis. Replace the seven screws (1) (0515-0372) and the six
screws (2) (0515-1946) that attach the fan bracket (3) to the
chassis. Torque all screws to 9 inch-pounds.
4. Replace the front frame. Refer to the “Front Frame Assembly” replacement
procedure.
5. Replace the instrument outer case. Refer to the “Instrument Outer Case”
replacement procedure.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9040B Option 508 and is the answer not in the manual?

Keysight N9040B Option 508 Specifications

General IconGeneral
BrandKeysight
ModelN9040B Option 508
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals