468 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Assembly Replacement Procedures
Input Connector
Replacement
1. Refer to Figure 15-113. Reinstall the input connector (1) into position in
the chassis. Replace the two screws (2) (0515-0372). Torque to
9 inch-pounds.
2. Reconnect the semi-rigid cable W2 to the input connector. Torque to
10 inch-pounds.
3. Replace the front frame. Refer to the “Front Frame Assembly” replacement
procedure.
4. Replace the instrument outer case. Refer to the “Instrument Outer Case”
replacement procedure.