Keysight Technologies N9040B UXA Signal Analyzer Service Guide 467
Assembly Replacement Procedures
Input Connector
Input Connector
Removal
1. Remove the instrument outer case. Refer to the “Instrument Outer Case”
removal procedure.
2. Remove the front frame. Refer to the “Front Frame Assembly” removal
procedure.
3. Refer to Figure 15-113. Disconnect the W2 semi-rigid cable from the input
connector (1).
Figure 15-113 Input Connector Removal
4. Remove the two screws (2) (0515-0372) that attach the input connector
to the chassis.