EasyManuals Logo

Keysight N9040B Option 508 User Manual

Keysight N9040B Option 508
542 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #472 background imageLoading...
Page #472 background image
472 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Assembly Replacement Procedures
Front Frame Assembly
Replacement
1. Refer to Figure 15-116. Carefully position the front frame assembly onto
the chassis. Ensure no cables are pinched. Ensure the O-ring is secured
over the RF connector. Using the T-20 driver, replace the fourteen screws
(1) (seven on each side) (0515-2044). Torque to 21 inch-pounds.
2. Using the T-10 driver, replace the one screw (2) (0515-0372). Torque to
9 inch-pounds.
3. Refer to Figure 15-115. Reattach the ribbon cable W1. Ensure the locking
tabs are engaged.
4. Refer to Figure 15-114. Reconnect the two cables (1) just behind the front
frame. Torque to 10 inch-pounds.
5. Replace the outer case. Refer to the “Instrument Outer Case” replacement
procedure.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9040B Option 508 and is the answer not in the manual?

Keysight N9040B Option 508 Specifications

General IconGeneral
BrandKeysight
ModelN9040B Option 508
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals