Keysight Technologies N9040B UXA Signal Analyzer Service Guide 117
Instrument Messages
Condition Messages
Condition Errors 67 to 95, Calibration Failure (Extended)
This series of errors corresponds to the bits in the
STATus:QUEStionable:CALibration:EXTended:FAILure sub-register. The second
column in the table shows the corresponding bit.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These error numbers can be seen in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
For example, error 71 indicates that the Characterize Preselector Failure has
been detected, error 1071 indicates that failure has been cleared.
This register is summarized as bit 9 of the STATus:QUEStionable:CALibration
register. See section “Condition Errors 36 to 64, Calibration Needed or Failed”
on page 114.
Err# Bit in
status
register
Message Error or
Warning
More Information
67 0 Align 9kHz to 30MHz
failed
W On PSA, this was error 13749
69 1 Align 30MHz to 1GHz
failed
W On PSA, this was error 13751
71 2 Characterize Preselector
failure
W The preselector characterization routine
failed.
73 3 unused
75 4 unused
77 5 unused
79 6 unused
81 7 unused
83 8 unused
85 9 unused
87 10 unused
89 11 unused
91 12 unused
93 13 unused
95 14 unused

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9040B Option 508 and is the answer not in the manual?

Keysight N9040B Option 508 Specifications

General IconGeneral
BrandKeysight
ModelN9040B Option 508
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals