Keysight Technologies N9040B UXA Signal Analyzer Service Guide 145
RF Section Troubleshooting (RF/Microwave Analyzers)
RF Section Description
This assembly contains the following circuits:
Input Low-pass filter (RF input signals < 3.6 GHz)
Optional Low Band Electronic Attenuator (0-24 dB attenuation control)
Optional Low Band Preamplifier and Limiter
RF Mixer #1 (RF input signals < 3.6 GHz)
1st LO Sub-system (3.8 GHz to 8.7 GHz)
1st IF Amplifier (5122.5 MHz)
Switched Filters (A13A2) provides wide BW band pass filter (low band)
LO and IM nulling (A13A2)
Mixer #2 (RF input signals 3.6 GHz to 13.599 GHz)
Microwave Input Amplifier
Mixer #3 (RF input signals 13.6 GHz to 26.5 GHz)
—LO Doubler
External Mixing Diplexer
A15 Front End Control Assembly
Provides the correct biasing and switch control logic to the following RF
assemblies:
Input Attenuator A
Input Attenuator B
—Low Band Switch
(Option LNP) Low Noise Path (standard on UXA)
YTF Preselector (Input signals > 3.6 GHz)
(Option MPB) Microwave Preselector Bypass (standard on UXA)
Front End Assembly including switched filter
—(Optional) High Band Preamplifier
(Optional) Low Band Preamplifier
(Option EA3) Electronic Attenuator (0-24 dB)
Refer to Chapter 5, “Front End Control Troubleshooting.” for detailed descriptions &
troubleshooting procedures