Keysight Technologies N9040B UXA Signal Analyzer Service Guide 271
Service and Diagnostics Menus
Service Key Descriptions
The Advisory Event “Alignment complete” is displayed when the
alignment is complete.
Align ADC can be interrupted by pressing the Cancel (ESC)
front-panel key. When this occurs, no new ADC alignment data will
be employed.
LO
Immediately executes an alignment of the LO subsystem. The
instrument will stop any measurement currently underway, perform
the alignment, then restart the measurement from the beginning.
A failure of LO will set the Error Condition “Align LO failed”. A failure
will not employ new LO alignment data.
Successful completion of LO will clear the Error Condition “Align LO
failed”.
The Advisory Event “Alignment complete” is displayed when the
alignment is complete.
Align LO can be interrupted by pressing the Cancel (ESC) front-panel
key. When this occurs, no new LO alignment data will be employed.
IF
Immediately executes an alignment of the IF subsystem. The
instrument will stop any measurement currently underway, perform
the alignment, then restart the measurement from the beginning.
A failure of IF will set the Error Condition “Align IF failed”. A failure will
not employ new IF alignment data.
Successful completion of IF will clear the Error Condition “Align IF
failed” and clear bit 6 in the Status.
The Advisory Event “Alignment complete” is displayed when the
alignment is complete.
Align IF can be interrupted by pressing the Cancel (ESC) front-panel
key. When this occurs, no new IF alignment data will be employed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9040B Option 508 and is the answer not in the manual?

Keysight N9040B Option 508 Specifications

General IconGeneral
BrandKeysight
ModelN9040B Option 508
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals