Keysight Technologies N9040B UXA Signal Analyzer Service Guide 275
Service and Diagnostics Menus
Service Key Descriptions
Temperature Gain & Tilt
This function controls both temperature adjustment mechanisms: overall
gain vs. temperature, and the temperature-proportional tilt of the gain vs.
frequency.
When Temperature Gain & Tilt is OFF, the Advisory Event “Flatness
Temperature Gain & Tilt OFF” will be displayed.
When Temperature Gain & Tilt is ON, the Advisory Event “Flatness
Temperature Gain & Tilt OFF” will be cleared.
IF Flatness
This function turns the corrections related to IF flatness On or Off.
When IF Flatness is OFF, the Advisory Event “IF Flatness corrections OFF”
will be displayed.
When IF Flatness is ON, the Advisory Event “IF Flatness corrections OFF”
will be cleared.
Snapshot Alignments On/Off
During normal auto alignment operation, the firmware will perform an RBW
and IQ alignment when the analyzer RBW state is changed, and every 10
minutes thereafter. These alignments can hinder troubleshooting the
analyzer since they can change correction factors when the state is
changed. Turning snapshot alignments off disables the RBW portion of
these alignments for more accurate troubleshooting. (Use IF Flatness
ON/OFF to control the IQ alignment.)
When Snapshot Alignments are OFF, the Advisory Event “Snapshot OFF”
will be displayed.
When Snapshot Alignments are ON, the Advisory Event “Snapshot OFF”
will be cleared.
LO Feedthru Nulling
Control is available to set the state of LO Feedthru nulling ON or OFF. In the
ON state, the software will determine when it is appropriate to null the LO
Feedthru. LO nulling is only available in low band.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9040B Option 508 and is the answer not in the manual?

Keysight N9040B Option 508 Specifications

General IconGeneral
BrandKeysight
ModelN9040B Option 508
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals