Keysight Technologies N9040B UXA Signal Analyzer Service Guide 355
Assembly Replacement Procedures
Top Brace and Card Cage Brace
Figure 15-4 Top Brace and Card Cage Brace Removal -
Serial Prefix < 5605 Options 508, 513, 526
Replacement
1. To replace the top brace or the card cage bracket, place them into the
correct position and attach the screws referred to in the removal process.
Torque to 9 inch-pounds.
2. Replace the instrument outer case. Refer to the “Instrument Outer Case”
replacement procedure.