EasyManua.ls Logo

Keysight N9040B Option 508 - Page 364

Keysight N9040B Option 508
542 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
364 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Assembly Replacement Procedures
RF Area - Options 508, 513, 526
Replacement
1. Refer to Figure 15-9 or Figure 15-10. Position the attenuator in the bracket
so that the ribbon connector end is “down”.
2. Replace the two screws (0515-0372) that attach the attenuator to the
bracket. Torque to 9 inch-pounds.
3. Position the attenuators and bracket into the chassis and attach with the
screws (3) (0515-0372). Torque to 9 inch-pounds.
4. Replace the ribbon cable and semi-rigid cables W2, W3, W4, and W51 to
the attenuators. Torque the semi-rigid cables to 10 inch-pounds.
5. Replace the RF bracket, front panel, and instrument outer case as
described on page 360.

Table of Contents

Related product manuals