EasyManua.ls Logo

Keysight N9040B Option 508 - Page 430

Keysight N9040B Option 508
542 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
430 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Assembly Replacement Procedures
A27 H1G Assembly (Option H1G)
Figure 15-78 Cable Routing
9. Locate the blue cable connected to J12 on the H1G assembly and install
the sma end through the rear panel hole marked IF 2 OUT. Secure with the
nut and washer removed earlier. Torque to 10 inch-pounds using a
5/16-inch wrench.
10.Re install the rear brace and rear panel using the screws removed earlier.
11.Locate the black plastic hole plug in the kit and install in the IF 2 IN hole
on the rear panel.
12.Re install the top brace and instrument enclosure.

Table of Contents