430 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Assembly Replacement Procedures
A27 H1G Assembly (Option H1G)
Figure 15-78 Cable Routing
9. Locate the blue cable connected to J12 on the H1G assembly and install
the sma end through the rear panel hole marked IF 2 OUT. Secure with the
nut and washer removed earlier. Torque to 10 inch-pounds using a
5/16-inch wrench.
10.Re install the rear brace and rear panel using the screws removed earlier.
11.Locate the black plastic hole plug in the kit and install in the IF 2 IN hole
on the rear panel.
12.Re install the top brace and instrument enclosure.