Keysight Technologies N9040B UXA Signal Analyzer Service Guide 525
Functional Tests
Frequency Response (Flatness), Preamp On
Figure 17-6 Frequency Response Test Setup
Power Sensor
(Opt P43, P44, P50)
Frequency Range: 50 MHz to 50 GHz
Amplitude Range: –65 dBm to –55 dBm
8487D
Power Splitter
(Opt P08, P13, P13)
Frequency Range: 50 MHz to 26.5 GHz
Tracking between ports: < 0.25 dB
11667B
Power Splitter
(Opt P43, P44, P50)
Frequency Range: 50 MHz to 50 GHz
Tracking between ports: < 0.40 dB
11667C
Synthesized Sweeper
Frequency Range: 50 MHz to 50 GHz
a
PSG
a. Maximum frequency of the sweeper must be equal to or greater than the maximum frequency of
the preamplifier option.
Item
Critical Specifications
(for this test)
Recommended
Keysight Model