Keysight Technologies N9040B UXA Signal Analyzer Service Guide            95
Instrument Messages
Event Messages
–221 Settings conflict; Must 
apply Amplitude Correction 
to make this unit available
These special units only apply when you are doing antenna 
measurements so you have to have a correction which 
includes Antenna Units enabled
–221 Settings conflict; No meas 
frequencies are above 3.6 
GHz
Optimize Preselector can only be performed on frequencies 
in high band i.e. freqs above 3.6 GHz. The current setup does 
not have input freqs (IF) in this range so an Optimize 
Preselector cannot be performed.
–221 Settings conflict; no source 
selected
You must select a source using Select Source before you can 
do this
–221 Settings conflict; Normal 
detector is not allowed with 
X scale is Log
The normal detector cannot be used when the x-axis scale is 
logarithmic.
Why not? Are the results weird? Frequency slewed?
–221 Settings conflict; Normalize 
is not available when Scale 
Type = Lin
Normalize does not support Linear amplitude scale, since 
the results are always presented as a dB ratio.
–221 Settings conflict; Normalize 
is not available while Demod 
View is on
The normalization (correction) function cannot be used if you 
are using the Demod View.
–221 Settings conflict; Normalize 
is not available while Trace 
Math is on
The Normalize function works by doing trace manipulation. 
So if trace math is on you cannot turn on normalization.
–221 Settings conflict; Only 
active Antenna Unit 
available; no other Y axis 
units
When a correction with antenna units is turned on, the only 
Y-Axis units you can have are those that match the Antenna 
Unit. Turn off the Correction or the Antenna Unit under 
Input/Output, Corrections
–221 Settings conflict; Option 
not available
You have attempted to perform an action for which a 
required option is not installed
–221 Settings conflict; Param 
only available when DUT is a 
freq converter
SCPI only message. The sideband and freq context 
parameters are only available when a freq conversion setup 
is in use. Change setup to contain a freq conversion to use 
these parameters.
–221 Settings conflict; Param 
only available when External 
LO Mode is Swept
SCPI only message. This parameter is only available when 
the LO mode is set to Swept. Change the LO Mode to Swept.
–221 Settings conflict; Param 
only available when External 
LO Mode is Fixed
SCPI only message. This parameter is only available when 
the LO mode is set to Fixed. Change the Freq Mode to Fixed.
–221 Settings conflict; Param 
only available when 
Frequency Mode is Fixed
SCPI only message. This parameter is only available when 
the Freq mode is set to Fixed. Change the Freq Mode to 
Fixed.
Err# Message Verbose/Correction Information