177
8. This screen allows you to set the Type of cabinet.
Leave this setting alone, and press [OK/MUTE].
If a Type C1, an additional screen will display to select
A-mode or B-mode, leave this selection unchanged,
and press [OK/MUTE].
9. This screen allows you to configure the cabinet for an
airflow sensor or not. Change the selection to AFS,
press [OK/MUTE].
10. This screen configures the unit for a UV light or not.
Leave this setting alone, and press [OK/MUTE].
11. This screen allows you to configure the cabinet for a
canopy or not. Leave this setting alone, and press
[OK/MUTE].
12. This screen allows you to configure the cabinet’s sash
height. Leave this setting alone, and press [OK/MUTE].
The unit will then return to the Home Screen in the new
configuration.
If a Type C1 in B-mode, an additional screen will be
displayed to select Active Protection Time. Leave this
setting alone, and press [OK/MUTE].