7-24
Using the Long-Term Chambers
second Deadband, Obs. Delay of 2 minutes, and Obs. Count of 3. This
protocol could then be repeated every hour for 240 hours (10 days). The
resulting data set would include 240 measurements, with each measurement
consisting of three 90-second observations.
6. Tap Start to begin the measurement.