2-22
UM-1151140EN, Rev. C, May 2016
Smart532
Laser
Safety
and
Regulatory
2.13.3 Recommended Separation Distances between Portable and Mobile RF Communications
Equipment and the Medical Equipment or Medical System
Recommended separation distances between portable and mobile RF communications equipment
and the M22 with ResurFX
The Smart532 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Smart532 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Smart532 as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01 0.116 0.116 0.233
0.1 0.369 0.369 0.738
1 1.167 1.167 2.334
10 3.689 3.689 7.378
100 11.667 11.667 23.334
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.