UM-1151140EN, Rev. C, May 2016
Laser
Safety
and
Regulatory
2.13.2 Electromagnetic Immunity (continued)
Guidance and Manufacturer's Declaration - Electromagnetic Immunity (continued)
The Smart532 is intended for use in the electromagnetic environment specified below. The customer or the
user of the Smart532 should assure that it is used in such an environment.
IMMUNITY
test
IEC 60601
TEST LEVEL
Compliance
level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms 150 kHz
to 80 MHz
3 V/m 80 MHz to
2,5 GHz
3 V
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the Smart532,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, should
be less than the compliance level in each frequency
range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
Smart532 is used exceeds the applicable RF compliance level above, the Smart532 should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as re-orienting or relocating the Smart532. b Over the frequency range 150 kHz to 80 MHz, field
strengths should be less than [V1] V/m.