2.13.2 Electromagnetic Immunity
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Smart532 is intended for use in the electromagnetic environment specified below. The customer or the
user of the Smart532 should assure that it is used in such an environment.
IMMUNITY Test
IEC 60601 test
level
Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst IEC
61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
N/A
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge IEC
61000-4-5
± 1 kV line(s) to
line(s) ± 2 kV line(s)
to earth
± 1 kV line(s) to
line(s) ± 2 kV line(s)
to earth
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
<5 % UT
(>95 % dip in UT) for
0,5 cycle
40 % UT (60 % dip
in UT) for 5 cycles
70 % UT (30 % dip
in UT) for 25 cycles
<5 % UT (>95 % dip
in UT) for 5 s
<5 % UT
(>95 % dip in UT) for
0,5 cycle
40 % UT (60 % dip
in UT) for 5 cycles
70 % UT (30 % dip
in UT) for 25 cycles
<5 % UT (>95 % dip
in UT) for 5 s
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
Smart532 requires continued
operation during power mains
interruptions, it is recommended that
the Smart532 be powered from an
uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
NOTE: Ut is the AC mains voltage prior to application of the test level.