Root NIBPT Chapter 3: Operation
www.masimo.com 59 Masimo
Systolic/Diastolic (SYS/DIA)
From the Systolic/Diastolic screen, touch Alarms, and then change any of the following
options:
Options Description
Alarm
Priority
Factory
Default
Settings
Configurable
Options
Systolic
High Limit
The High Limit is upper
threshold that triggers an
alarm.
Medium 220
42-259 in steps of 1,
or Off
When set to Off,
alarm is disabled
Systolic Low
Limit
Low Limit is the lower
threshold that triggers an
alarm.
Medium 75
41-258 in steps of 1,
or Off
When set to Off,
alarm is disabled
Diastolic
High Limit
The High Limit is upper
threshold that triggers an
alarm.
Medium 110
22-199 in steps of 1,
or Off
When set to Off,
alarm is disabled
Diastolic
Low Limit
Low Limit is the lower
threshold that triggers an
alarm.
Medium 35
21-198 in steps of 1,
or Off
When set to Off,
alarm is disabled