A - 2 VM600 MPS hardware manual (standard version) MAMPS-HW/E
Edition 17 - February 2018
Overall
ENVIRONMENTAL SPECIFICATIONS
Power supply perturbations:
• AC line voltages
(AC-input version
of the RPS6U)
115 V
AC
±20%.
230 V
AC
+15%, î›20%.
• AC line frequencies 50/60 Hz
• AC line harmonics
2
nd
=4.5%, 3
rd
= 4.5%, 4
th
=3.0%,
5
th
= 1.5%, 6
th
= 1.5%, 7
th
=3.0%
• DC line voltage variations
(DC-input versions
of the RPS6U)
+30%, î›25%
Electromagnetic compatibility (EMC):
• Emission According to IEC/EN 61000-6-4
• Immunity According to IEC/EN 61000-6-2
Electrical safety Conforms to the following electrical safety
standards:
IEC/EN 61010-1: Safety requirements for electrical
equipment for measurement, control and laboratory
use
Altitude Max. 2000 m (6560 ft).
Note: Reduced air density affects cooling ability.