Operating Instructions - CPC6050 45
Modular Pressure Controller
CPC6050
6.4.3.4 Rate Setpoint
The Rate Setpoint (SVR Module only) button allows the user to set the rate of pressure change when the
CPC6050 is controlling up or down to a setpoint (gure 6.4.3.4). The rate is limited from 0.1% to 20% of
span of the primary transducer / second of the active channel.
Figure 6.4.3.4 - Rate Setpoint
6.4.3.5 Stability Parameters
Stability parameters for the controlled pressure can be congured using the Stable Window and Stable
Delay buttons (SVR Module and Pump Module). When the controller enters a stable condition the pres-
sure indication color on the Home App will change from white to green. The Stable Window button allows
the user to enter a value as a percentage of the highest range transducer. This value represents the pres-
sure window within which any setpoint value would be considered stable by the user. The Stable Delay
button lets the user add a desired delay until the pressure value is considered stable while being in the
stable window.
Stability
Parameters
Figure 6.4.3.5 - Stability Parameters