EasyManua.ls Logo

mensor CPC6050 - GE PACE (SCPI) Emulation Command Set

mensor CPC6050
145 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instructions - CPC6050 97
Modular Pressure Controller
CPC6050
7.10 GE PACE (SCPI) Emulation Command Set
The CPC6050 will accept and respond to the SCPI remote commands of GE PACE 5000/6000 remote
commands using the option SCPI_GE. Section 11, Technical Assistance explains the process of en-
abling the SCPI GE command set. Table 7.10.1 shows a complete list of supported commands.
7.10.1 SCPI GE Supported Commands and Queries
Table 7.10.1 SCPI GE Supported Commands and Queries
Command Response/Function
CALCulate
[:LIMit]
:LOWer<n> Set the minimum control limit *
:LOWer? Set the minimum control limit *
:UPPer<n> Set the maximum control limit *
:UPPer? Set the maximum control limit *
CALibration
[:PRESsure][R]
:POINt? Returns the number of calibration points used in
the active sensor’s linearity correction. Calibra-
tion functionality must be enabled by sending the
calibration password. See SYST:PASS:CEN
:ZERO
:AUTO Start autozero process
:AUTO? Returns 1 if autozero process is running, 0 other-
wise
VALVe ON | OFF | 1 | 0 ON or 1 puts in vent mode, OFF or 0 puts in mea-
sure mode
:VALVe? Returns 1 if in vent mode, 0 otherwise
INPut
:LOGic? Returns a single digit number that is a mask
of the digital input states, i.e. a number of 7 in
binary is 111 which indicates all digital inputs are
on. Then current pressure reading.
Ex// INP:LOG?
:INP:LOG 7, 14.50021
:STATe? Returns a single digit number that is a mask
of the digital input states, i.e. a number of 7 in
binary is 111 which indicates all digital inputs are
on.
Ex// INP:LOG:STAT?
:INP:LOG :STAT 0

Table of Contents

Related product manuals