Operating Instructions - CPC6050 53
Modular Pressure Controller
CPC6050
6.4.5.1 Remote Command Set
The remote command set parameter provides a choice of the Mensor command set, the WIKA SCPI or
DPI5xx command set (Figure 6.4.5.1 - A). All sets of commands are listed in Section 7, Remote Opera-
tion.
Figure 6.4.5.1 - A - Remote Command Set
The DPI5xx command also provides an option to set the Checksum (Figure 6.4.5.1 - B) which when set
to ON or AUTO appends a mod base 100 checksum to end of any remote communication response.
A proper mod base 100 checksum must be sent to the instrument when the checksum is ON. Sending
the checksum is optional when set to AUTO. The ow control is set to NONE as default internally. The
termination character sets the output termination character for all remote communication command sets
(Figure 6.4.5.1 - C).
Figure 6.4.5.1 - B - Checksum Figure 6.4.5.1 - C - Termination character