Operating Instructions - CPC6050 53
Modular Pressure Controller
CPC6050
6.4.5.1 Remote Command Set
The remote command set parameter provides a choice of the Mensor command set, the WIKA SCPI or
DPI5xx command set (Figure 6.4.5.1 - A). All sets of commands are listed in Section 7, Remote Opera-
tion.
Figure 6.4.5.1 - A - Remote Command Set
The DPI5xx command also provides an option to set the Checksum (Figure 6.4.5.1 - B) which when set
to ON or AUTO appends a mod base 100 checksum to end of any remote communication response.
A proper mod base 100 checksum must be sent to the instrument when the checksum is ON. Sending
the checksum is optional when set to AUTO. The ow control is set to NONE as default internally. The
termination character sets the output termination character for all remote communication command sets
(Figure 6.4.5.1 - C).
Figure 6.4.5.1 - B - Checksum Figure 6.4.5.1 - C - Termination character