72 Operating Instructions - CPC6050
Modular Pressure Controller
CPC6050
7.6 Serial
Set the Serial communication parameters as shown in Section 6.4.5 Remote Application. The serial com-
munication port allows the CPC6050 to communicate in RS-232 format with computers, terminals, PDAs,
or similar hosts.
These parameters should be set to match your host computer. Default settings are: 9600 baud, 8 data
bits, 1 stop bit, no parity, and no echo.
If echo is ON, the CPC6050 will immediately echo back characters sent over the serial port. The Serial
function allows the user to set the RS-232 serial port settings by selecting from the choices provided:
Baud
9600
19200
38400
57600
115200
Data Bits
7
8
Stop Bits
1
2
Parity
Even
Odd
None
Echo settings
On
O
7.6.1 Serial Cable Requirements
RS-232 communications are transmitted over a three conductor, shielded cable terminated in a standard
DB9 connector on the instrument end, and a dierent gender connector on the host end. The proper pin-
outs are shown in the following illustration.