92 Operating Instructions - CPC6050
Modular Pressure Controller
CPC6050
[:AMPLitude]? Returns the setpoint
:SLEW <n> Sets the rate setpoint
:SLEW? Returns the rate setpoint
:TOLerance? Returns the stable window
:TOLerance<n> Sets the stable window
CALCulate
:LIMit
:LOWer<n> Set the minimum control limit
:LOWer? Set the minimum control limit
:UPPer<n> Set the maximum control limit
:UPPer? Set the maximum control limit
:SYSTem
:DETECT SLOW | FAST | CANCEL Not used, kept for backwards compatibility
:DETECT? Returns “0”
7.8.2 Error Messages
All remote commands sent to the CPC6050 are shown in the Trouble Shooting App, under the “Remote”
button. If there is a syntax error, an error message will appear below the errant command. Local errors
and remote command errors will appear under the “Error” button. Maximum of 100 errors are stored and
can be retrieved.
The error messages can also be seen remotely when “SYST:ERR?” is sent any error in the error queue is
returned, otherwise the following is returned: 0,“No Error”.
Error String Returned
0, “No error”
1,”Overow”
102, “Syntax error”
114,”Parameter Out Of Range”
500,”<s>” (where <s> is the undened error error string generated by the pneumatic module)
602,”Sensor not available”
999,”<s>” (where <s> is the undened (unique) error string)