EasyManuals Logo

mensor CPC6050 User Manual

mensor CPC6050
145 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #98 background imageLoading...
Page #98 background image
98 Operating Instructions - CPC6050
Modular Pressure Controller
CPC6050
INSTrument
:CATalog[R]? Returns the sensor pressure maximum in native
units, native pressure units, native pressure type.
Ex// INST:CAT?
:INST:CAT “1000.000psig”
:CATalog
:ALL? Returns the primary sensor’s pressure maximum
in native units, native pressure units, native pres-
sure type, secondary sensor’s pressure maxi-
mum in native units, native pressure units, native
pressure type, Barometer. If a sensor does not
exist, it is not included.
Ex// INST:CAT:ALL?
:INST:CAT “1000.000psig”, “300.0000psig”,
BAROMETER
:LIMits? Returns the instrument’s pressure maximum in
native units, native pressure units, native pres-
sure type, current upper control limit, current
lower control limit.
Ex// INST:LIM?
:INST:LIM “1000.000psig”, 1000.000,
-15.0000
:MACaddress? Returns the ethernet interface medium access
control (MAC) hardware address in XX-XX-XX-
XX-XX-XX format. Any letters will be upper case.
Ex// :INST:MAC?”
:INST:MAC ”84-EB-18-B2-97-AB”
:SENSe[R]? Returns the sensor pressure maximum in native
units, native pressure units, native pressure type.
Ex// INST:SENS?
:INST:SENS “1000.000psig”
:SENSe[R]
:CALD? Returns the sensor calibration date in the format
YYYY, MM, DD
:FULLscale? Returns the sensor maximum pressure, minimum
pressure in active units
:NEGCal? Returns 1. Negative calibration support is always
available
:SN? Returns the sensor serial number