Operating Instructions - CPC6050 89
Modular Pressure Controller
CPC6050
:RATE[R]? Returns the rate/sec from range R
:BAROmetric? Returns the barometric pressure
CALibration
[:PRESsure][R]
:MODE? Returns 1=calibrated or 0=not calibrated
:DATE? Returns date of cal “MM/DD/YY”
:DATE <i,i,i> Sets date of cal YYYY,MM,DD
:ZERO? Returns zero oset
:ZERO <n> Sets the zero oset
:ZERO:RUN Start Autozero sequence
:ZERO:STOP Stop Autozero sequence
:ZERO:INITiate? Returns zero status
:ZERO:INITiate Ignored
SENSe
[:PRESsure][R]
:NAME? Returns transducer name string
:MODE? Returns “ABSOLUTE” or “GAUGE”
:MODE ABS | GAUGE Sets pressure type
:ABS? Returns native transducer type
0=GAUGE 1=ABSOLUTE
:RESolution? Returns resolution (oat)
:RANGe
[:UPPer]? Returns maximum range
:LOWer? Returns minimum range
:UNIT
[:NAME]? Returns ASCII units (mixed case)
:VALue? Returns the units conversion factor
:REFerence
[:HEIGht] <n> Sets the head pressure height
:HEIGht? Returns the head pressure height
:MODE? Returns “OFF”, “GAS”, or “LIQUID”
:MODE OFF | GAS | LIQUID Sets the head pressure mode
:MEDium<n> Sets the medium density
:MEDium? Returns medium
:ACTive <n> Sets the active transducer
ACTive? Returns the active transducer
SYSTe m
:DATE <i,i,i> Not used, kept for backwards compatibility
:DATE? Not used, does not cause an error, does not
return a response