Probes and Biopsy
18-50
Trade Name Chemical Name Procedures
Minncare
®
Cold
Sterilant
22% Hydrogen
Peroxide
4.5% Peroxyacetic
Acid
Dilute the sterilant with sterilized purified
water (1:20).
Immersed time: 11 hours.
Temperature: 20
℃-25℃.
Please refer to the instructions provided by
the manufacturer of the solution for details.
Before safety and performance is affected, plastic bracket NGB-006, NGB-007, NGB-008
and NGB-012 can be sterilized by Minncare COLD STERILANT for at least 245 times (11
hours for one time).
îš„ STERRAD 100S low-temperature hydrogen peroxide gas plasma sterilization
system
Chemical
name
Trade name Procedures
Hydrogen
peroxide gas
plasma
Hydrogen
peroxide vapor
Please refer to the instructions provided by the
producer of the solution for details.
z Refer to the instruction of STERRAD 100S sterilizing system provided by the
manufacturer for operation instructions and cautions.
z The STERRAD 100S low-temperature hydrogen peroxide gas plasma sterilization
system is available for metal needle-guided brackets.
îš„ High-pressure steam sterilization (only applicable for metal guided-bracket)
Autoclaving (moist heat) 121
° C for 20 minutes.
NOTE: 1. Repeated sterilization may degrade the safety and performance
of the needle-guided bracket.
2. The high-pressure steam/ immersion sterilization do not affect the
bracket dual life, and the dual life is affected by the daily
application of the bracket. Please check the appearance of the
bracket before using.
18.2.9 Storage and Transportation
1. Don’t use the carrying case for storing the needle-guided bracket. If the carrying case
is used for storage, it may become a source of infection.
2. Between examinations, keep the needle-guided bracket in a sterile environment.
3. When the needle-guided bracket is sent to your MINDRAY representative for repair,
be sure to disinfect or sterilize it and keep it in the carrying case to prevent infection.
4. Sterilize the carrying case as necessary.
5. Store or transport the needle-guided bracket under the following ambient conditions:
z Ambient temperature: -20°C to 55°C
z Relative humidity: 30% to 85% (no condensation)