Electrical Safety Inspection
D-3
D.5 Earth Leakage Test
Run an Earth Leakage test on the device being tested before performing any other
leakage tests.
The following outlet conditions apply when performing the Earth Leakage test.
z normal polarity( Normal Condition);
z reverse polarity( Normal Condition);
z normal polarity with open neutral(Single Fault Condition);
z reverse polarity with open neutral(Single Fault Condition).
îš„ LIMITS
For UL60601-1,
z 300 μA in Normal Condition.
z 1000 μA in Single Fault Condition.
For IEC60601-1,
z 500 μA in Normal Condition.
z 1000 μA in Single Fault Condition.
D.6 Enclosure Leakage Test
The following outlet conditions apply when performing the Enclosure Leakage test.
z normal polarity( Normal Condition);
z reverse polarity( Normal Condition);
z normal polarity with open neutral(Single Fault Condition);
z reverse polarity with open neutral(Single Fault Condition);
z normal polarity with open earth(Single Fault Condition);
z reverse polarity with open earth(Single Fault Condition).
îš„ LIMITS
For UL60601-1,
z 100μA in Normal Condition.
z 300 μA in Single Fault Condition.
For IEC60601-1:
z 100μA in Normal Condition.
z 500 μA in Single Fault Condition.
D.7 Patient Leakage Current
Patient leakage currents are measured between a selected applied part and mains earth.
All measurements have a true RMS only.
The following outlet conditions apply when performing the Patient Leakage Current test.
z normal polarity( Normal Condition);
z reverse polarity( Normal Condition),
z normal polarity with open neutral(Single Fault Condition);
z reverse polarity with open neutral(Single Fault Condition).
z normal polarity with open earth(Single Fault Condition);
z reverse polarity with open earth(Single Fault Condition).