87
D F−97 Prohibition of calibration
Initial value=00000 00000=Possible to calibrate
00001=Prohibition from calibration
D F−98 ZERO clear
Zero compensated data by zero set function can be cancelled.
When the
key is pressed with “F−98” displayed, “ZCLR” can be displayed.
(At the same time, the display lights on and off.)
At this point, press the
key when suspending the ZERO clear is desired.
Measurement mode can be returned and Zero clear will not be executed.
When the
key is pressed while “ZCLR” display lights on and off, “F−98” display can
be returned. Now, ZERO clear has completed.
D F−99 Memory clear
Setting from F−01 to F−97 recorded at EEPROM will return to the default value.
When the
key is pressed with the display of F−99, then “FCLR” display can be
obtained. (At the same time, the display lights on and off.) At this point, press the
key when suspending memory clear is desired. Measurement mode can be returned and
Memory clear will not be executed.
When the key is pressed during “FCLR” load display lights on and off, and after
about 1 second, it will become “F−00” display and the operation of Memory clear has
completed.
Warning ● Never use the following functions because they may destroy the
functions at internal of the instrument.
F−00 F−02 F−05 F−07 F−13 F−14 F−18 F−19
F−23 F−24 F−25 F−26 F−27 F−28 F−29 F−33
F−37 F−38 F−39 F−46 F−47 F−48 F−49 F−58
F−59 F−60 F−61 F−62 F−63 F−64 F−65 F−66
F−67 F−68 F−69 F−72 F−73 F−74 F−75 F−76
F−77 F−78 F−79 F−80 F−81 F−82 F−83 F−84
F−85 F−86 F−87 F−89 F−95 F−96