D F−55 Setting the terminator of RS−232C/RS−422
(Effective when the option is installed.)
Initial value=00001 00000:CR
00001:CR+LF
D F−56 Setting the decimal point of sending code of RS−232C/RS−422
(Effective when the option is installed.)
Initial value=00000 00000:No decimal point
00001:Yes of Decimal point
D F−57 Setting the ID number of RS−422(Effective when the option is installed.)
Initial value=00000 Setting range:00000〜00031
D F−70 Changing Peak function / A/Z function
Initial value=00000 00000:Effective in the Peak function
00001:Effective in the A/Z function
D F−71 Setting the area of ZERO, A/Z and stored data
Initial value=00001 0=RAM
1=EEPROM
10
1
digit:A/Z data
10
0
digit:ZERO data
※Selects whether the data at the time of ZERO A/Z will be recorded at the EEPROM or
not. In case that the high frequency of use of ZERO, A/Z and A/Z OFF, the durability
times will be over, so set “0”. The durability times for the EEPROM will be approx. one
million times.
D F−88 Selection of serial interface operation mode
(Effective when the option is installed.)
Initial value=00000 00000=Stream
00001=Stream+Automatic print
00002=Stream+Manual print
D F−90 Increment value (For reference)
※The increment value set when the calibration is applied can be referred to.
※The setting can’t be changed in the function.
D F−91 The maximum display value (For reference)
※The maximum display value set when the calibration is applied can be referred to.
※The setting can’t be changed in the function.
D F−92 The actual load value (For reference)
※The actual load value set when the calibration is applied(LCAL) can be referred to.
※When the calibration except LCAL is made, this value will not change.
※The setting can’t be changed in the function.
D F−93 Zero calibration value (For reference)
※The input voltage value that has read as the initial load value at the time of executing
calibration can be referred to.
※The setting can’t be changed in the function.
D F−94 Span calibration value (For reference)
※The input voltage value at the time of the maximum display can be referred to.
※The setting can’t be changed in the function.