85
D F−42 Setting the polarity of BCD output (Effective when the option is installed.)
Initial value=00000 00000=Negative logic
00001=Positive logic
D F−43 Setting the logic of BCD flag output (Effective when the option is installed.)
Initial value=00000 00000=Negative logic
00001=Positive logic
D F−44 Setting the logic of BCD P.C. output (Effective when the option is installed.)
Initial value=00000 00000=Negative logic
00001=Positive logic
D F−45 Setting the BCD P.C. width(Effective when the option is installed.)
Initial value=00000 00000=125 ms
00001=25 ms
D F−50 Setting the operation mode of RS−232C
(Effective when the option is installed.)
Initial value=00001 00000=Stream mode
00001=Command mode
D F−51 Setting the target of output at the time of stream mode of RS−232C
(Effective when the option is installed.)
Initial value=00000 00000=TRACK/Gross weight
00001=PEAK/Net weight
※In the Command mode, the data are transferred from the host to the command.
In the Stream mode, the latest data will be output continuously.
D F−52 Setting the baud rate of RS−232C/RS−422
(Effective when 3the option is installed.)
Initial value=00003 00000=1 200 bps
00001=2 400 bps
00002=4 800 bps
00003=9 600 bps
00004=19 200 bps
D F−53 Setting the data bit length and parity of RS−232C/RS−422
(Effective when the option is installed.)
Initial value=00021 10
0
digit:Setting data bit length
0=8 bit
1=7 bit
10
1
digit:Setting parity
0=No parity
1=Even number parity
2=Odd number parity
D F−54 Setting the stop bit of RS−232C/RS−422
(Effective when the option is installed.)
Initial value=00000 00000:1 bit
00001:2 bit