84
D F−20 Setting the target of analog output
Initial value=00000 00000:TRACK/Gross weight
00001:PEAK/Net weight
D F−21 Display value at the time of the minimum analog output
Initial value=00000 Setting rage −99999〜99999
D F−22 Display value at the time of the maximum analog output
Initial value=02000 Setting rage −99999〜99999
D F−30 Setting comparator operation
Initial value=00011 0=OFF
1=ON
10
0
digit:Comparator S1
10
1
digit:Comparator S2
D F−31 Setting the target of comparator
Initial value=00000 0=TRACK/Gross weight
1=Synchronous with display
10
0
digit:Comparator S1
10
1
digit:Comparator S2
D F−32 Setting the direction of comparator
Initial value=00000 0=Greater or equal
1=Less or equal
10
0
digit:Comparator S1
10
1
digit:Comparator S2
D F−34 Setting the condition of Hysteresis operation for comparator
Initial value=00001 00000=ON delay
00001=OFF delay
D F−35 Setting the hysteresis data width for comparator
Initial value=00000 00000=Hysteresis data width OFF
Setting rage:00000〜00099
Unit:1D
Data width of 99D at the setting of “00099”.
D F−36 Setting Hysteresis time width for comparator
Initial value=00000 00000=Hysteresis time width OFF
Setting rage:00000〜00099
Unit:0.1 s
Data width of 9.9 s at the setting of “00099”
D F−40 Setting the target of BCD output(Effective when the option is installed.)
Initial value=00000 00000=TRACK/Gross weight
00001=PEAK/Net weight
D F−41 Setting the logic of BCD output (Effective when the option is installed.)
Initial value=00000 00000=Negative logic
00001=Positive logic

Table of Contents