D F09 Setting zero tracking time width
Initial value00020 00000Zero tracking OFF
Setting range0000000099
Unit0.1 s
Time width of 9.9 s with the setting “00099”.
Effective only when the value of 0000100099 is set with the F08.
D F10 Setting of the target of HOLD
Initial value01111 10
0
digitLoad display
10
1
digitComparator S1, S2 contact output, LED display
10
2
digitAnalog output
10
3
digitOptional output
Out of the target at the setting “0”, and the target of HOLD is available at the setting
“1”.
D F11 Setting the CHECK value
Initial value00004 Setting range0000000014
Unit0.2 mV/V(Setting can be made by every 0.2 mV/V)
Approx. 1.4 mV/V CHECK value at the setting “00014”.
D F12 Setting bridge power supply voltage
Initial value00000 000005V
000012.5 V
D F15 Setting digital filter for stabilized filter
Initial value00003 00000Stabilized filter OFF
Setting range0000000005
Average of 2
m
times
Selects average times of digital filter for stabilized filter.
When the figure grows larger, the filter will be stronger, then the effects from
vibrations and so on are scarcely shown on the display.
Effective only when setting is made with the value from 00001 to 00999 with the
F16, and the value from 00001 to 00999 is set with the F17.
D F16 Setting time width for stabilized filter
Initial value00020 00000Stabilized filter OFF
Setting rage0000000999
Unit0.01 s
Time width of 9.99 s at the setting of “00999”.
Effective only when the value from 00001 to 00005 with the F15 and the value from
00001 to 00999 with the F97 are set.
D F17 Setting data width for stabilized filter
Initial value00020 00000Stabilized filter OFF
Setting rage0000000999
Unit1D
Data width of 999D at the setting of “00999”.
Effective only when the value from 00001 to 00005 with the F15 and the value from
00001 to 00999 with the F16 are set.