126
1
Remove the connecting cable at the current output terminal.
2
Measure the voltage between the current output terminal + and .
Set the connecting range to DCmA range for the measuring instrument such as
tester and so on.
3
Apply operational check on current output in the Check mode.
(Refer to the paragraph 718.)
Load resistance is
260 kΩ or less.
1
Put the connecting cable for the current output terminal where it was.
2
Check the setting related with current output.
F20 Selection of target for the analog output.
F21 Display value at the time of the minimum analog output.
F22 Display value at the time of the maximum analog output.
(Refer to the paragraph 715)
Contact with Minebea.
Load resistance should be 260 kΩ
or less.
Set correctly. Refer to the
paragraph 715, 8 and 91.
Set the HOLD input open.
Inform Minebea about the contents of
failure and situation at site in details.
10
NO
YES
Varies to 4 mA, 12
mA and 20 mA.
NO
YES
Setting is correct.
NO
YES
External control
input or HOLD is applied.
NO
YES