21
Procedures
Press the key.
The load display will show “2000”, and the digit
of minimum display will flash on and off.
When the calibration has completed already, the
registered value of maximum display at that time
will be displayed.
6
By the setting of minimum scale, the digit
of minimum display that flashes on and off
will be as follows:
The minimum scale 1, 2, 5 10
0
digit
The minimum scale 10 10
1
digit
Set the maximum display value with the right
keys. Setting range for the maximum display
value will be (the minimum scale×100)〜99 990.
In order to make effective use of the performance,
set within the following ranges.
When setting is made over the range as below,
there may have a possibility of unstable display
and so on.
key :Set value carry key
key :Set value inclement key
key :Set value initialization key
Setting range for the
maximum display value
100〜10 000
200〜20 000
500〜50 000
1 000〜99 990
The minimum scale
1
2
5
10
It can be increased continuously by the
continuity press of the key.
7
Press the key.
The load display will show “S MV”.