If an error occurs midway during execution of the key sequence (amplitude
compression error “15 Amplitude compression failure” excepted, however),
the execution terminates at that point.
If an overload occurs midway during execution of the key sequence, the
execution is terminated only when either one of [SWEEP STOP] or
[OSCOFF] in the menu [Input][ACTION] is [ON].
If a key sequence involving a REPEAT measurement is executed, the
REPEAT measurement is repeated permanently without going to the next
execution. Therefore, do not record/store/execute any key sequence
involving REPEAT measurement. Press the GPIB LOCAL key to terminate
the REPEAT measurement in the key sequence.
When any access, control or inquiry is made from the GPIB (i.e., when the
status turns to REMOTE) during execution of key sequence, the execution is
terminated.
During execution of key sequence operations, all key entries are ignored
except for the GPIB LOCAL key.
If the work area memory cannot be secured at the time of starting execution,
the execution does not start with an error message (“Memory overflow”)
displayed.