4.18 Memory
FRA5087
4-44
4.18 Memory
The FRA5087 is equipped with memories which are to be used for recording/storing measured
data.
Types
Battery back-up provision
Memory capacity
Mass memory
No (deleted with power off)
No less than 20,000 points*1
Permanent memory
Yes
No less than 2,000 points
Equalizing memory
No (deleted with power off)
No less than 20,000 points*1
*1: The total capacity of the mass and equalizing memories is no less than 20,000 points of
measurement data.
Use the menu [Memory] for recording, storage or deletion to/of memory contents.
[STORAGE]
To store the currently displayed data into memory. Use the function key to select the
location of storage.
[MASS STORAGE] : To store in the mass memory
[PERMANENT STORAGE] : To store in the permanent memory
[EQL STORAGE] : To store in the equalizing memory.
The Equalizing function (menu [Measure][BASIC
FUNCTION][EQUALIZING]) becomes valid, only
when data are stored in this memory (i.e., equalizing
memory).
[DELETE MASS DATA]
[DELETE PERMANENT DATA]
To delete the stored content of the master data memory or of the permanent data memory.
A list of stored data content will be displayed; move the cursor to the desired data using the
ITEM
key and then press the function key [DELETE] to delete the desired data.
Functions of Function keys:
[PREV] : To go back to the previous page
[NEXT] : To go to the next page
[DELETE] : To execute deletion
CAUTION
If you press the function key [DELETE CUR.TAG] while the [MEMORY]
window is displayed, the current display is deleted. Once the display has
been deleted, it cannot be restored.