3.2 Display at power “ON” and initial settings
FRA5087
3-15
Table 3-1 List of initialization values (continued)
Setting items
Initialization
values
At
shipment
from
factory
At data
destruction
At
initialization
executed
Notes/Remarks
OTHERS
TITLE SET
(Cleared)
Also cleared at
re-measurement
and at power
supply switch on
BUZZER
ON
DATE SET
YEAR
1988
The actual
shipment date is
set at shipment
from factory
MONTH
1
DAY
1
TIME SET
HOUT
0
The actual
shipment time is
set at shipment
from factory
MINITE
0
3.2.3 Warm-up
It takes more than 30 minutes after the power supply switching-on for the internal temperature of
the FRA5087 to reach stable.
Perform measurement right after calibration is made, which shall be made after sufficient time to
warm-up has been made. Note that the measurement accuracy specification is met under the
condition immediately after calibration.
Conduct re-calibration when environment temperature has been changed.