5.5 Other settings
WF1946B 5-23
③ Set the pulse width to 0.5 ms (
ENTRY
→
WIDTH
→
・
→
5
→
m
).
④ Set oscillation mode.
(
MODE
→
BURST
→TYPE = TRIG, SOURCE = EXT , DELAY = 0.1 ms,
MARK = 1.0 cycle, STOP-LEVEL = ON −100%)
⑤ Apply the trigger signal to the TRIG/SWEEP IN connector.
Other:
• Double pulse output: At above step ④, set MARK = 2.0 cycles.
• Manual trigger: At above step ④, set SOURCE = EXT
and press the
MAN TRIG
key.
(Do not connect anything to the TRIG/SWEEP IN connector.)