EasyManua.ls Logo

NF WF1946B - Operation Error

NF WF1946B
204 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6.1 Error message
WF1946B 6-3
Operation error
Error message Cause or corrective measures
DATA OUT OF RANGE
Input values outside of permissible setting range.
Confirm permissible setting range and again input.
Upper and lower limits can be easily checked by using the
MODIFY dial to change the values.
INVALID NUMERIC DATA
Keypad input data invalid (e.g., decimal point only).
SETTINGS CONFLICT 001
Start or stop value set to 0 during LOG sweep and sweep cannot
be executed.
SETTINGS CONFLICT 002
Combined center and span exceed the permissible sweep type
(e.g., frequency) setting range.
SETTINGS CONFLICT 003
Combined modulation type (e.g., frequency) and either Deviation
or depth exceed the permissible modulation type setting range and
modulation cannot be executed.
SETTINGS CONFLICT 004
Combined modulation type (e.g., frequency) and either Deviation
or depth exceed the permissible modulation type setting range.
SETTINGS CONFLICT 007
Combined period and pulse width exceed the permissible duty
setting range.
SETTINGS CONFLICT 008
Combined frequency (FREQ) and frequency difference (FREQ)
or frequency ratio (RATIO) settings exceed the permissible
frequency setting range.
SETTINGS CONFLICT 010
Since the mode is SWEEP or MODU for
(variable duty),
DUTY-VALID cannot be set.
continued next page

Table of Contents