7.3 Performance tests
7-6 WF1946B
Amplitude accuracy
Connection: Use coaxial cable to connect FUNCTION OUT to AC voltmeter 1.
Setting: After setting initialize, set the amplitude, output range and waveform as indicated in
the following table.
Measurement: Measure the true rms output voltage for each waveform. (CH1,CH2)
Judgment: The normal ranges are indicated in the table.
Waveform Output range Setting Normal range
10 V 20 Vp-p/open (7.071 Vrms/open) 7.004 to 7.138 Vrms
, ,
10 V 20 Vp-p/open ( 5.774 Vrms/open) 5.719 to 5.828 Vrms
10 V 20 Vp-p/open ( 10.00 Vrms/open) 9.905 to 10.095 Vrms
10 V 10 Vp-p/open ( 3.536 Vrms/open) 3.493 to 3.578 Vrms
10 V 5 Vp-p/open (1.768 Vrms/open) 1.738 to 1.798 Vrms
10 V 2 Vp-p/open (0.707 Vrms/open) 0.684 to 0.730 Vrms
1 V 2 Vp-p/open (0.7071 Vrms/open) 0.699 to 0.716 Vrms
DC offset accuracy
Connection: Connect FUNCTION OUT to a DC voltmeter.
Setting: After setting initialize, set the DC mode, output range and DC offset as indicated in
the following table.
Measurement: Measure the output voltage. (CH1,CH2)
Judgment: The normal ranges are indicated in the table.
Output range DC offset setting Normal range
10 V ±10.000 V/open ±9.880 to ±10.12 V
10 V ±5.000 V/open ±4.905 to ±5.095 V
10 V ±2.000 V/open ±1.920 to ±2.080 V
10 V ±1.000 V/open ±0.925 to ±1.075 V
10 V 0.000 V/open −0.070 to +0.070 V
1 V ±1.0000 V/open ±0.985 to ±1.015 V
1 V 0.0000 V/open −0.010 to +0.010 V