4.3 Sorting the Part (Comparator)
4-10 ZM2371/ZM2372
The display format for the lower/upper limit values and reference values, and unit
The lower limit value, upper limit value and reference values are construed according to the
type of display parameter or the display format.
For example, “F” is used as the unit of lower/upper limit value when displaying/sorting the
capacitance by the absolute value or deviation, meanwhile “%” is used as the unit when
displaying/sorting the capacitance by the deviation%
The following table shows the case when the lower limit value=1, upper limit value=2 and
reference value=3 are set during sorting the inductance (L).
Deviation is 1H
(equivalent to 4H for original
measured value)
Deviation is 2H
(equivalent to 5H for original
measured value)
Deviation is 1%
(equivalent to 3.03H for
original measured value)
Deviation is 2%
(equivalent to 3.06H for
original measured value)
Initializing the setting of comparator
To newly set the upper/lower limit value of comparator, it will be convenient to initialize the
setting at first.
To initialize the upper/lower limit value, select the CLEAR in the comparator setting
menu, and the following comparator initializing menu is displayed.
Comparator CLEAR
1)Execute Option
Done Completion message
Perform the EXIT
operation to return to one-previous menu.
Press the
_
1
_
key to excute the initialization, and the completion message will be displayed
for a short time, and then one-previous menu will come back.
Table 4–1 Initialization contents of comparator
Comparator function
(Bin sorting)
All upper limit values
(Value, ON/OFF)
Sorting for BIN2 to BIN14
All lower limit values
(Value, ON/OFF)
Comparison for secondary
parameter
The display format and comparison format are not initialized.