LIFEPAK 500 automated external defibrillator Battery Maintenance
10-7
Battery Maintenance Page 6 of 12
Recharging a
Rechargeable
Battery Pak
(continued)
WARNING!
Possible loss of power during patient care. Using an improperly
maintained battery to power the LIFEPAK 500 AED may cause premature
power loss. Use only the LIFEPAK 500 AED battery charger to charge the
LIFEPAK 500 rechargeable battery pak.
A fully charged battery is not harmed if it remains connected to the battery
charger. However, if a battery is disconnected and then reconnected, the battery
charger begins the 10 hours of fast charge again. Additional battery fast-charge
cycles, without discharging, can reduce battery life.