LIFEPAK 500 automated external defibrillator Replacement Procedures
11-66
Replacing Wave Shaping Inductor—Monophasic AED Page 2 of 3
To install the new Wave Shaping Inductor:
1. Slide a new inductor tie-wrap through the attachment point in the inductor
cavity as shown at left.
2. Connect the P13(+) spade-lug to the Energy Storage Capacitor according to
the detail at left.
3. Slide the replacement inductor into the inductor cavity in the lower enclosure,
with the inductor wires to the left, P14 towards the rear, and P13(+) towards
the front. Continue with the steps for routing, dressing, and tie-wrapping the
wiring on the next page.
Note: Be sure to position tie-wrap buckles as shown on the next page, tighten,
and trim excess. The inductor tie-wrap must be pushed to the rear of the case.
This allows for maximum clearance between the tie-wrap and the PCB when
installed.
Inductor
Tie-Wrap
Wave
Shaping
Inductor
Inductor
Cavity
To Relay at
P14
To Defib
Capacitor
P13 (+)