LIFEPAK 500 automated external defibrillator Replacement Procedures
11-17
Replacing Case Page 1 of 2
Application The case replacement procedure is a composite of procedures for installing a new
case and other assemblies as necessary. If you are opening the case simply for
access to internal components, jump to Removing Top Cover.
Preparation Before replacing the case, first follow these procedures:
■ Removing Top Cover
WARNINGS!
Shock hazard. Servicing of this AED must be performed by properly trained
individuals. This AED may retain potentially lethal charges accessible
inside the AED at any time–even when off. Follow procedures carefully for
discharging the Energy Storage Capacitor and disabling Auto Battery Test.
Shock hazard. The Energy Storage Capacitor carries high voltage.
Remove the battery and discharge the capacitor before handling.
■ Discharging Energy Storage Capacitor
CAUTION!
Possible component damage. The Main PCB assembly contains static
sensitive devices (SSDs). To avoid damage, observe the special handling
practices described in the section titled Electrostatic-Sensitive Device
Handling.