LIFEPAK 500 automated external defibrillator Replacement Procedures
11-42
Removing Main PCB—Biphasic AED Page 1 of 3
Application The following procedure describes removing a Main PCB that you intend to
reinstall later. If you intend to replace the Main PCB of a LIFEPAK 500 Biphasic
AED with a new one, perform the Replacing Main PCB—Biphasic AED
procedure.
Preparation Before removing the Main PCB, you must first follow these procedures:
Removing Top Cover
WARNING!
Shock hazard. The Energy Storage Capacitor carries high voltage.
Remove the battery and discharge the capacitor before handling. See:
Discharging Energy Storage Capacitor
To remove the Main PCB:
1. Place the AED on the work surface with the handle facing forward.
2. Disconnect P9 of the speaker cable from J9 on the right underside of the
Main PCB.
3. LIFEPAK 500 Biphasic AED equipped with Readiness Display only:
Disconnect P20 of the Readiness Display cable from J20 of the Main PCB.