LIFEPAK 500 automated external defibrillator Battery Maintenance
10-15
AED Cleaning Page 2 of 3
External Cleaning
Procedures
Clean the exterior of the LIFEPAK 500 AED by wiping the surface with any of the
following solutions:
Soap and water
Quaternary ammonium compounds
Isopropyl alcohol
Peroxide (peracetic acid) solutions
WARNING!
Shock or fire hazard. Do not immerse or soak any portion of this AED in
water or any other fluid. Avoid spilling any fluid on the AED or accessories.
CAUTION!
Possible case damage. Do not use bleach, bleach dilution, or phenolic
compounds. Do not steam or gas sterilize. Do not autoclave this AED.